Model 600 Barocel ® Absolute Vacuum Pressure Transducer


Model 600 Barocel ® Absolute Vacuum Pressure Transducer

The model 600 Barocel Transducer is an advanced design capacitance manometer which uses patented technology to improve zero stability after temperature excursions or pressure overloads. An integral vacuum reference of all-welded construction contains material to ensure long term stability.

In vacuum systems, the model 600 is the idea replacement for McLeod, Pirani, or thermocouple gauges. Since pressure is sensed by electrically measuring pressure induced displacement of a thin metal membrane, calibration accuracy is unaffected by the chemical composition of input media.

Model 600 Barocel ® Absolute Vacuum Pressure Transducer

Features and Benefits

  • Wide range
    4 decades.
  • High accuracy – 0.15% of reading
  • High resolution – 0.01% full scale
  • Unaffected by overpressure – repeats within 0.01% full scale after being exposed to 2.4 bar (35 psia)
  • True total pressure – reading independent of gas composition
  • High level output – 0.10 volts d.c, full scale, linear with pressure
  • Superior zero stability – patented design eliminates errors due to temperature and overpressure effects
  • Stable zero reference
  • Alternative for McLeod, Thermocouple and Pirani gauges
  • System versatility – compatible with a variety of readouts and controllers
  • Corrosion resistant – measuring cell constructed of Inconel/Monel for high corrosion resistance

Edwards Model 600 Barocel ® Absolute Vacuum Pressure Transducer

Model 600 Barocel ® Absolute Vacuum Pressure Transducer: Technical specification


Pressure ranges:
0-10, 0-100, 0-1000 mbar
0-10, 0-100, 0-1000 Torr
Ranges to 10000 Torr also available
Accuracy*: ± ( 0.5% of reading + 0.015 of full scale)
0.5% of reading also available
Resolution: Better than ±0.01% full scale
Operating temperature: 0 – 65 °C
Temperature affect on zero: < 0.005% full scale/ °C
Temperature affect on sensitivity < 0.02% reading/ °C
Overpressure without damage: 1.25 x full scale range or
2.4 bar (35psia) absolute exposure
Material exposed to process: Inconel and Monel
Internal Volume: 8.3cm3 (includes 1/2” dia tube)
Time constant: 8 msec
Output signal: 0-10 V d.c. Linear with pressure
Power requirements: ±15 V d.c. ±5% regulated to ±15 at 30mA
Weight: 0.7 KG (1.5lb)
Standards
Electronic design: EN 61010-1
Electromagnetic compatibility: EN 61326 Industrial Location, Class A Emissions


Model 600 Barocel ® Absolute Vacuum Pressure Transducer: Ordering Information

Edwards Model 600 Barocel ® Absolute Vacu

Model 600 Barocel ® Absolute Vacuum Pressure Transducer: Accessories

Adapter cable Burocel 600 to AGC (female FCC68) 0.5m: D40003050